To save to your PC, right mouse-click on the file name and select 'Save target As'.
PDF documents need the Acrobat(R) Reader from Adobe(R): Adobe, Acrobat and the Acrobat Logo are trademarks of Adobe Systems Incorporated. Free download from www.adobe.com
TSI's VelociCalc Air Velocity Meters are easy to use from the first time you pick them up. The meters measure velocity and temperature, and with a few simple keystrokes the calculate flowrate, perform multi-value averaging, and determine minimum and maximum readings.
The VelociCalc Features
- Extended velocity range of 0 to 30 m/s
- Temperature range of -17.8 to 93.3°C
- Easy-to-read digital display
- Variable time constant modes available for a steady display when measuring fluctuating flows
- 94.0 cm telescoping probe with etched length marks to make duct traverse measurements easier
- Sampling function allows for easy recording of multiple measuring points
- Statistics function can display average, maximum and minimum values, and the number of recorded samples
- Flowrate feature allows for simple and quick calculations of volumetric flowrate when the user inputs the duct shape and size
- Optional portable printer provides a hard copy documentation of your readings
- Includes a battery check that monitors battery level to ensure accurate readings
- NIST*-traceable calibration certificate
MODEL OPTIONS
The Models 8345 and 8346 both measure air velocity, temperature, calculate flowrate, perform multi-value averaging, and determine minimum and maximum readings. The Model 8346 has the added feature of an articulating probe for measurements in ceiling outlet flows or clean benches.
APPLICATIONS
- HVAC duct measurements
- Fume hood face velocity tests
- Clean room studies
- Wind tunnel work
- Filter face velocity measurements
- Indoor Air Quality tests
SPECIFICATIONS
VELOCITY (ALL MODELS):
Range:- 0 to 30 m/s (0 to 6,000 ft/min)
Accuracy1&2:- 3.0% of reading or ±0.15 m/s (±3 ft/min) whichever is greater
TEMPERATURE (all models):
Range:- -17.8 to 93.3°C (0 to 200°F)
Resolution:- 0.1°C (0.1°F)
Accuracy3:- ±0.3°C (±0.5°F)
INSTRUMENT TEMPERATURE RANGE (all models):
Operating (Electronics):- 5 to 45°C (40 to 113°F)
Operating (Probe):- -17.8 to 93.3°C (0 to 200°F)
Storage:- -30 to 90°C (-22 to 194°F)
VOLUMETRIC FLOWRATE (all models):
Range4:- 0.1 to 195,000 l/s, 0.0424 to 702,000 m3/hr, 0.2 to 2,700,000 ft3/min
DUCT SIZE (all models):
Range: - 1 to 100 cm in increments of 0.5 cm, 100 to 255 cm in increments of 1 cm
(1 to 100 inches in increments of 0.5 inches, 100 to 255 inches in increments of 1”)
AVERAGING CAPABILITY (all models):
Range: - Up to 255 values each of velocity, temperature
TIME CONSTANT (all models):
Range:- Adjustable from 1 to 20 seconds
RESPONSE TIME (all models):
To velocity :- 200 msec
To temperature :- 8 seconds
PRINTER INTERFACE:
Type: - Serial
Baud Rate:- 1200
*U.S. National Institute of Standards and Technology
1 Temperature compensated over an air temperature range of 5 to 65°C (40 to 150°F).
2 The accuracy statement of 3.0% of reading or ±3 ft (±0.15 m/s), applies to 30 ft/min through 6,000 ft/min.
3 Accuracy with instrument case at 25°C (77°F), add uncertainty of 0.03°C/°C (0.05°F/°F) for change in instrument temperature. Add uncertainty of 0.5°C (0.9°F) for exposure to RF radiation levels of 3 V/m per IEC 801-3.
4 Actual range is a function of maximum velocity and duct size.
Range
Resolution
0 to 30 m/s
0.01 m/s
-17.8 to 93.3°c
0.1 °c
BSRIA Instrument Solutions reserve the right to change Instrument
specifications without notification. Hire Instrument Specifications may
differ from those claimed by the manufacturer.