BSRIA Instrument Solutions - a comprehensive service for your Test and Measurement requirements
BSRIA Instrument Solutions
Old Bracknell Lane West, Bracknell, Berkshire, RG12 7AH, UK
Tel: +44 (0) 1344 459314 Fax: +44 (0) 1344 465556
Sales : sales@bis.fm Hire : hire@bis.fm Calibration : calibration@bis.fm
Return to home page Contact Us View contents of your Quote
 Calibration / Multifunction Electrical Calibrators
New search

Fluke : 725



Downloads
Download   Fluke 725 Multifunction Process Calibrator PDF file 141KB
To save to your PC, right mouse-click on the file name and select 'Save target As'.
PDF documents need the Acrobat(R) Reader from Adobe(R):
Adobe, Acrobat and the Acrobat Logo are trademarks of Adobe Systems Incorporated.
Free download from www.adobe.com


Take along the new Fluke 725 Multifunction Process Calibrator, and you're equipped to test and calibrate almost any process parameter. Measure and source mA, volts, temperature (RTDs and thermocouples), frequency, ohms, and pressure, using optional pressure modules.

Do you need to calibrate transmitters? The 725's split display lets you view input and output values simultaneously. For valve and I/P tests, you can source mA while measuring pressure. The 725 has auto-stepping and auto-ramping for remote testing, plus 25% stepping for fast linearity tests.

From the moment you pick it up, the Fluke 725 is ready to perform. Its simple controls, without menus, make operation easy. Its memory functions make set up fast. And its rugged design lets it work as hard as you do.

The Fluke 725 Multifunction Process Calibrator is a powerful, yet easy-to-use field calibrator. Use the measure and source functions to test and calibrate almost any process parameter.

• Small, streamlined shape makes it easy to carry
• Rugged, reliable design stands up to field use
• Easy to read measure/source screen lets you view input and output simultaneously
• Measure volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors and transmitters
• Source/simulate volts, mA, thermocouples , RTDs, frequency, ohms, and pressure to calibrate transmitters
• Measure/source pressure using any of 29 Fluke 700Pxx Pressure Modules
• Source mA with simultaneous pressure measurement to conduct valve and I/P tests
• Support flow meter testing with frequency and CPM functions
• Perform fast linearity tests with auto- step and auto-ramp features
• Power transmitters during test using loop supply with simultaneous mA measurement
• Handling of fast pulsed RTD transmitters and PLCs, with pulses as short as 10 ms
• Store frequently used test setups for later use

MEASUREMENT ACCURACY
========================

VOLTAGE DC
30.000 V 0.02%+ 2 counts (upper display)
30.000 V 0.02%+ 2 counts (lower display)
100.00 mV 0.02%+ 2 counts
-10.00 mV to 75.00 mV 0.025 % + 1 count (via TC connector)

CURRENT DC
24.000 mA 0.02%+ 2 counts

RESISTANCE
0.0 to 400.0 ohms 0.1ohms (4-wire), 0.15ohms (2- and 3-wire)
401 to 1500 ohms 0.5ohms (4-wire), 1ohms (2- and 3-wire)
1500 to 3200 ohms 1ohms (4-wire), 1.5ohms (2- and 3-wire)

FREQUENCY
2.0 to 1000.0 CPM 0.05 % + 1 count
1.0 to 1100.0 Hz 0.05 % + 1 count
1.00 to 10.00 kHz 0.05 % + 1 count

SENSITIVITY
1 V peak-to-peak-minimum
Pressure Accuracy from 0.025% of range using any of 29 pressure modules.
(for detailed specifications refer to pressure modules in options and accessories)
Modules available for differential, gage, vacuum, absolute, dual and high pressure.

SOURCE ACCURACY
========================

VOLTAGE DC
100.00 mV 0.02%+2 counts
10.000 V 0.02%+2 counts
-10.00 mV to 75.00 mV 0.025 % + 1 count (via TC connector)

CURRENT DC
24.000 mA (Source) 0.02%+ 2 counts
24.000 mA (Simulate) 0.02%+ 2 counts

RESISTANCE
15.0 to 400.0 ohms 0.15 ohms (exc. current 0.15 to 0.5 mA), 0.1 ohms (exc. current 0.5 to 2 mA)
401 to 1500 ohms 0.5 ohms (excitation current 0.05 to 0.8 mA)
1500 to 3200 ohms 1 ohms (excitation current 0.05 to 0.4 mA)

FREQUENCY
2.0 to 1000.0 CPM 0.05%
1.0 to 1100.0 Hz 0.05%
1.00 to 10.00 kHz 0.25%
Waveform: 5 V p-p squarewave, -0.1 V offset

RTDS AND THERMOCOUPLES

MEASURE ACCURACY:
NI-120 0.2 °C
PT-100 (385) 0.33 °C
PT-100 (393) 0.3 °C
PT-100 (JIS) 0.3 °C
PT-200 (385) 0.2 °C
PT-500 (385) 0.3 °C
PT-1000 (385) 0.2 °C
Resolution: 0.1 °C
J: 0.7 °C
K: 0.8 °C
T: 0.8 °C
E: 0.7 °C
R: 1.8 °C
S: 1.5 °C
B: 1.4 °C
L: 0.7 °C
U: 0.75 °C
N: 0.9 °C

RESOLUTION
J, K, T, E, L, N, U: 0.1 °C, 0.1 °F
B, R, S: 1 °C, 1 °F

SOURCE ACCURACY:
NI-120 0.2 °C
PT-100 (385) 0.33 °C
PT-100 (393) 0.3 °C
PT-100 (JIS) 0.3 °C
PT-200 (385) 0.2 °C
PT-500 (385) 0.3 °C
PT-1000 (385) 0.2 °C

RESOLUTION: 0.1 °C
Note: Accuracy stated for 4-wire measurement.
J: 0.7 °C
K: 0.8 °C
T: 0.8 °C
E: 0.7 °C
R: 1.4 °C
S: 1.5 °C
B: 1.4 °C
L: 0.7 °C
U: 0.75 °C
N: 0.9 °C
Resolution J, K, T, E, L, N, U: 0.1 °C, B, R, S: 1 °C

TECHNICAL DATA
=========================
Maximum current: 22 mA, short circuit protected
Step Functions Source functions: Voltage, current, resistance, frequency, temperature
Steps 25% of range, 100% of range

ENVIRONMENTAL SPECIFICATIONS
Operating Temperature -10 °C to +55 °C
Storage Temperature -20 °C to 71 °C
Operating Altitude 3000 m

SAFETY SPECIFICATIONS
Agency Approvals EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992

MECHANICAL & GENERAL SPECIFICATIONS
Size 130 x 236 x 61 mm
Weight 0.65 kg
Batteries 4 AA alkaline batteries
Battery Replacement Separate battery compartment, accessible without breaking calibration seal
Side port connections Pressure module connector, also used for remote real-time programming






BSRIA Instrument Solutions reserve the right to change Instrument specifications without notification. Hire Instrument Specifications may differ from those claimed by the manufacturer.


Models, availability and pricing

Fluke 725
Availability Price
Hire Call
Sale Call
Calibration Call
Get quote to...   quantity...

 

Email this page

Do you know a colleague who will be interested in these products? Send them a link to this page.

 

 

 

 

Top of page

Home : Sales : Hire : Services : Calibration : Contact Us