BSRIA Instrument Solutions - a comprehensive service for your Test and Measurement requirements
BSRIA Instrument Solutions
Old Bracknell Lane West, Bracknell, Berkshire, RG12 7AH, UK
Tel: +44 (0) 1344 459314 Fax: +44 (0) 1344 465556
Sales : sales@bis.fm Hire : hire@bis.fm Calibration : calibration@bis.fm
Return to home page Contact Us View contents of your Quote
 Calibration / Pressure Calibrators
New search

Druck : DPI610/IS & 615/IS






These self-contained, battery powered packages contain a pressure generator, fine pressure control, device energising (not IS version) and output measurement capabilities, as well as facilities for 4 to 20 mA loop testing and data storage. The rugged weatherproof design is styled such that the pressure pump can be operated and test leads connected without compromising the visibility of the large dual parameter display. The mA step and ramp outputs and a built-in continuity tester extend the capabilities to include the commissioning and maintenance of control loops.

The technically advanced Druck DPI 610 and DPI 615 portable calibrators are the culmination of many years of field experience with the company’s DPI 600 series.

These self-contained, battery powered packages contain a pressure generator, fine pressure control, device energising (not IS version) and output measurement capabilities, as well as facilities for 4 to 20 mA loop testing and data storage. The rugged weatherproof design is styled such that the pressure pump can be operated and test leads connected without compromising the visibility of the large dual parameter display. The mA step and ramp outputs and a built-in continuity tester extend the capabilities to include the commissioning and maintenance of control loops.

Intrinsically Safe versions are available as complete maintenance tools and portable calibration standards for pressure instruments and control loops in hazardous ’Zoned‘ areas. Certified to BASEEFA and CENELEC standards, the DPI 610IS and DPI 615IS can reduce response times to breakdowns and emergencies by removing the need for ’Hot Permits‘ and gas detection equipment. This gives peace of mind to all those responsible for safety within hazardous areas.

A highly accurate and easy to use calibrator is only part of the solution for improving overall data quality and working efficiency. The DPI 610 and DPI 615, with data storage and RS 232 interface, reduce calibration times and eliminate data recording errors. The DPI 615 also provides error analysis for field reporting of calibration errors and pass/fail status. In addition, procedures downloaded from a PC automatically configure the DPI 615 to pre-defined calibration and test routines.

IMPROVED PERFORMANCE
---------------------
The DPI 610/615 series combine practical design with state of the art performance, summarised as follows:

Accuracy: 0.025% FS for ranges 70mbar to 700mbar
Ranges: 2.5 mbar to 700 bar including gauge, absolute and differential versions

Integral pneumatic pressure source Models
–850 mbar to 20 bar

Integral hydraulic pressure source Models: 0 to 400 bar

Measure: Pressure, mA, V, switch state (open/closed) and ambient temperature.

Output: Pressure, mA step, mA ramp, mA value Energising supplies: 10 and 24 Vdc (not IS version)

Data storage: 10,000 values ( In 999 chunks )

Documenting (DPI 615 only): Error analysis with pass/fail status and graphs. Two way PC communication for transferring procedures and results.

Remote pressure sensors: Up to 10 digitally characterised sensors per calibrator.

Please note that not all ranges/models are available for rental. Please contact our Hire department for availability.

SIMPLIFIED OPERATION
--------------------------
Druck’s knowledge of customer needs, combined with innovative design, results in high performance, multi-functional calibrators which are simple to use. The key to simple operation is the Task Menu.

Specific operating modes such as P-I, switch test and leak test are configured at the touch of a button by menu selection. Featuring highly reliable pneumatic and hydraulic assemblies and self-test routines, the DPI 610/615 series can be relied upon time and time again for field calibration in the most extreme conditions.






BSRIA Instrument Solutions reserve the right to change Instrument specifications without notification. Hire Instrument Specifications may differ from those claimed by the manufacturer.


Models, availability and pricing

DPI610/IS & 615/IS
Availability Price
Hire Call
Sale Call
Calibration Call
Get quote to...   quantity...

 

Email this page

Do you know a colleague who will be interested in these products? Send them a link to this page.

 

 

 

 

Top of page

Home : Sales : Hire : Services : Calibration : Contact Us

Privacy policy : Terms and Conditions : Logout :