Airflow Developments : Telescopic Pitot Tube Max Length 0.98m x 4mm dia




Downloads
Download   Pitot Tubes Tele PDF file 100KB
To save to your PC, right mouse-click on the file name and select 'Save target As'.
PDF documents need the Acrobat(R) Reader from Adobe(R):
Adobe, Acrobat and the Acrobat Logo are trademarks of Adobe Systems Incorporated.
Free download from www.adobe.com


The Pitot static tube is a universally accepted means of establishing air velocity in ducts not fitted with permanent flow monitoring devices. When used with a high quality manometer the accuracy and dependability of results is far greater than with any non-fundamental method.


An important advantage of the telescopic Pitot static tube is that it is only necessary to drill a few small access holes in the duct at the point where the flow measurement is to be made, and this feature makes the method the unquestioned choice for site work, where time and access is at a premium and a large number of in-duct measurements may have to be made at different points in the system.

By comparison with some other velocity indicating devices, the Pitot static tube and manometer has a further advantage in that no flow of air passes through the instrument. This eliminates worries about pressure drop through the tubes to the indicating instrument, and for practical purposes there is no limit to the distance between the Pitot tube and the manometer.

The shape of the nose, the size, number and position of the pressure holes etc, are all critical in determining the characteristics of the finished Pitot tube and the modified ellipsoidal nose form used in Airflow Developments' tubes is recommended in BS ISO 3966:2008. This design was originally evolved by the U.K. National Physical Laboratory and is outstanding for its combination of high accuracy with insensitivity to errors in alignment. The calibration factor is very close to unity (0.997).

TELESCOPIC PITOT STATIC TUBE ADVANTAGE:

FEATURES

·Durable and robust for on site conditions up to 100°C.
·Air velocity/temperature calculator supplied.
·Telescopic probe with range 200 to 980mm long.
·Carrying case supplied.

BENEFITS

·Compatible with all manometers
·Compact probe with maximum portability and high accuracy.
·Telescopic sections can be used as insertion depth markers.
·No limit to the distance between the Pitot tube and the manometer.

APPLICATIONS

·HVAC commissioning.
·Process Control lines.
·Research testing.
·Heat exchanger balancing.
·Combustion efficiency tests.

SPECIFICATION

Minimum overall length (compressed) 200mm
Maximum overall length (extended) 980mm
Maximum tube diameter 13mm
Overall head length 64mm
Head diameter 4mm
Maximum working temperature 100°C






BSRIA Instrument Solutions reserve the right to change Instrument specifications without notification. Hire Instrument Specifications may differ from those claimed by the manufacturer.


Models, availability and pricing

Max Length 0.98m x 4mm dia
Availability Price
Hire Call
Sale Call
Calibration Call
Get quote to...   quantity...

 

Email this page

Do you know a colleague who will be interested in these products? Send them a link to this page.

 

 

 

 

Top of page

Home : Sales : Hire : Services : Calibration : Contact Us

Privacy policy : Terms and Conditions : Logout :